Plastic tanks and piping are particularly
susceptible to grounding problems, because of poor earth grounding.
Dirty motors from electrically driven pumps or mixers, other
electrical equipment and even conductivity probes leak voltage into
the solution which feeds through the pH electrode and electronic
circuit. This leads to invalid readings and shortened
electrode life.
The DA series electrode combats these problems
by utilizing a solution ground pin and internal battery powered
ground interrupt circuit. The battery has an estimated 2 year
service life. The DA series electrodes are part of our Flat
Surface, self cleaning series quick disconnect designs which
offer all the benefits of our time tested and proven self cleaning
series and more!
The classical diagnostic symptom is when the
system calibrates and works fine outside of the tank or line, but
drifts or reads incorrectly when installed in the system.
The battery powered
internal amplifier permits long cable runs to be used between the
electrode and the transmitter-controller, even 1,000 feet or more
!
These cost effective electrodes are mounted into standard
Sensorex gland fittings and can be used for submersion installations
as well as in 3/4", 1" and 2" size pipe tee's. They use our
standard cap cables for ease of use. Special Fluoride
Resistant versions are also available.
|