JENCO
- 9250
|

$350.00 |
Dissolved Oxygen handheld meter, uses
polarographic type sensor with replaceable
membranes and programmable pressure
compensation, 600 to 1000 mBar and salinity
compensation, 0 to 40.0 ppt. Features large
multi function LCD display for D.O. and
temperature and an RS-232C interface. ATC is a
10K thermistor. Power 115/230V AC or battery.
Download
Spec Sheet
Manual
|
ACCESSORIES AND KITS
|
\KITS
# |
DESCRIPTION |
PRICE |
|
9250 YKA |
Kit (9250Y:meter w/batteries, and
900YDOL-10:DO LAB probe, 9005YDOK-L:membrane kit,
AC adapter, 9250F case |
$725.00 |
|
9250 YKB |
Kit (9250Y:meter w/batteries and
900YDOL-10:DO, LAB probe, 9005 YDOK-L membrane kit
|
$695.00 |
FEATURES
- High Reliable Mechanical Touch Keys With
Tactile & Audio Feedback
- Displays Concentration To 0.1% Mg/L (Ppm)
- Pressure & Salinity Compensation
- RS-232 Output
- CE Approved
- Battery /AC Adaptor
- Power Down Memory Backup
- Displays Percent Saturation To 0.1%
- Splash Resistant/Watertight Case
|
SPECIFICATIONS
|
0 to 20.00 |
0.01 |
± 0.2% of spa |
|
0 to 199.9 |
0.1 |
± 0.2% of span |
|
-6.0 to 46.0 |
0.1 |
± 0.1% ± 1 digit |
GENERAL
|
Communication output |
RS-232C |
|
Memory backup |
Yes |
|
Temp. compensation |
Thermistor,10K Ohms at 25 Centigrade
|
|
Display :dissolved oxygen:temp. |
15 mm high:8.7 mm high LCD
|
|
Power supply:batteries: adaptor |
0 to 50 Centigrade
|
|
Battery life (alkaline battery) |
Six AAA size batteries: 9 V ac
adaptor |
|
Dimensions (W x L x H) |
83 mm X 222mm X 721mm
|
|
Weight |
410 gm (batteries included)
|
|